The liquid exhaust is microcomputer controlled for linear and consistent liquid cool down, programmable within a specified range. Typical applications include wrapped and unwrapped porous and non-porous hard goods, gowns or towel packs and liquids in self-venting or unsealed containers. The selectable temperature range is from 230 F to 275 F (110 C to 135 C) and from 219 F to 275 F (104 C to 135 C) for liquid cycles. APPLICATION For general-purpose gravity or vacuum steam sterilization of hospital instruments and supplies. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters. Custom cycle names can be designated for each cycle and each cycle can be reconfigured for easy access. Up to 17 cycles can be easily accessed in two easy steps. 1 CUSTOMER: REFERENCE: 833HC-E SERIES VACUUM/GRAVITY STEAM STERILIZERS FOR HEALTHCARE APPLICATIONS PRODUCT SPECIFICATION PRODUCT The Model 833HC-E Vacuum/Gravity Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal.